Error Codes | Suggestions |
108 SOCKET_VOLTAGE_FRONT_SHELL_ERROR | Voltage detected on the front jaws when meter is not energized. Check test setup. |
118 SOCKET_AMP_FAULT | Socket is unable to read the
proper current on front jaws. Check meter insertion. Reboot the socket with the meter attached. |
111
SOCKET_LOAD_CONFIG_ERROR | Error reading the EEPROM of the socket. Reverts to default values. |
112
SOCKET_CONFIG_CHECKSUM_ERROR | Error reading the EEPROM of the
socket. Reverts to default values. |
120
SOCKET_LINE_VOLTAGE_OUT_RANGE | Line voltage is outside of
socket parameters (between 100 VAC and 480 VAC). Check input voltage. |
121
SOCKET_LINE_FREQUENCY_OUT_RANGE | Line frequency is outside of
socket parameters (between 58 Hz and 60 Hz). Check input frequency. |
203
OPTDEV_NOT_DETECTED | No optical device
detected. Connect optical device to IO port. |
402
METER_CONFIG_CHECKSUM_ERROR | Connect optical device to IO port. |
405
METER_VOLTAGE_OUT_OF_RANGE | Meter voltage is outside of
socket parameters (between 100 VAC and 480 VAC). Check input voltage. |
406
METER_FREQUENCY_OUT_OF_RANGE | Meter frequency is outside of
socket parameters (betwenn 58 Hz and 60 Hz). Check input frequency. |
407
METER_NOT_FOUND | No meter detected. Check meter insertion. |
408
METER_NONE_INSTALLED | No meter installed during KH
autodetect. Check meter insertion. |
502
CALIBRATION_VOLTAGE_OUT_OF_RANGE | Voltage on front jaws is out of
range during calibration (within 5% of selected nominal value). Check input voltage. |
503
CALIBRATION_CURRENT_OUT_OF_RANGE | Current on front jaws is out of
range during calibration (within 5% of nominal value). Check test setup. |
504
CALIBRATION_FREQUENCY_OUT_OF_RANGE | Frequency on front jaws is out
of range during calibration (within 5% of 60 Hz). Check input frequency. |
505
CALIBRATION_IFB_SCALE_OUT_OF_RANGE | Current on front jaws is out of
range. Check test setup. |
506
CALIBRATION_BURDEN_TOO_HIGH | Cannot reach maximal calibration
current value. Check test setup.. |
507
CALIBRATION_NO_PULSE | No pulse is received on the
socket IO port. Check test setup. |
1009
QUESTIONS_CHECKSUM_ERROR | Error reading the EEPROM of the
socket. Reverts to default values. |
1111
REMOTE_CONFIG_CHECKSUM_ERROR | Error reading the EEPROM of the
socket. Reverts to default values. |
1216
REPORTS_CHECKSUM_BAD | Error reading one or multiple
reports in the EEPROM of the socket. Resets affected report slots. |
1217
REPORTS_ALL_SLOTS_USED | No more available report save
slots. Transfer test reports from socket to PC and delete reports from socket. |
1302
SYSTEM_INIT_ERROR | Could not initialize
system. Please contact Probewell Lab Inc Technical Service. |